The word that best defines and determines its characteristics is, without doubt, universal. It has multiple rotors with a wide range of reducers, which enables it to work with volumes from 0.2 ml. to 100 ml., with angle fixed and swing out options. This equipment covers all types of needs of the users and offers a range of medium speeds for routine processes. To optimise the performance of the equipment in this range, we have increased its capacity with a swing out rotor for up to 28 x 15 ml. tubes and an angle fixed rotor for 32 to 15 ml./15 ml. conical tubes.
It has a control that defines the functions beyond the operation values, thus becoming a piece of equipment perfectly integrated into your processes. Versatile, functional,simple, indispensable in your lab.
Features
Maximize
Minimize
LED screen:
- Shows RPM/ RCF and time.
- Speed programming in 50 RPM/ 10 xg steps.
- Timer countdown/up from “0” or at “set RPM/ RCF” for reproducible tests.
- Timer from 1 to 99 min. programmable in 1 min. steps and hold position.
- Acceleration and deceleration control in 3 steps: fast, soft and free.
- Acoustic and visual messages on screen showing the user the status of the equipment.
Easy to use
Maximize
Minimize
- Microprocessor controlled.
- Induction motor maintenance free (brushless).
- Rotors list in memory.
- Listado de rotores en memoria.
- Noise level: below 60 dB.
- Start, stop, open lid and short spin with adjustable speed buttons.
- Option of free/locked adjustment of RPM/ RCF along the run.
- Automatic lid opening, programmable.
- Low height for easy access.
- Last values remain in memory.
- Over-speed protection.
Safety
Maximize
Minimize
- Lid provided with security systems:
– Automatic lid lock system, motorized with double lock.
– Emergency lid-lock release.
– Lid dropping protection.
– Port in the lid for calibration and operation checking.
- Unbalance detection and switch off.
- Protection safety ring between the centrifugation chamber and the housing.
- Chamber of centrifugation in stainless steel (easy cleaning).
- Buckets can be removed with the lid closed.Hermetic lids
- Rotors and adapters autoclavable, easy to install by the user.
- Forced ventilation to reduce temperature increasing
- Automatic disconnection for energy saving, with deactivation option.
Normas y directivas comunitarias
Maximize
Minimize
Eu Directives: 2011/65/EU, 2012/19/EU, 2014/30/EU, 2014/35/EU, 98/79/EC.
Standars: EN 61010-1, EN 61010-2-101, EN 61010-2-020, EN 61326-2-6, EN 61326-1.
Versions
Maximize
Minimize
Code | Dimensions (mm) (w x d x h) | Net weight (Kg) | Voltage (V) | Frequency (Hz) | Consumption (W) |
---|---|---|---|---|---|
CE 126 | 410x520x320 | 36 | 220-230 | 50-60 | 280 |
CE 127 | 410x520x320 | 36 | 110-120 | 50-60 | 280 |